Search found 75 matches

by Pico Stuart
Fri Apr 09, 2010 1:37 pm
Forum: PicoScope 6 for Windows
Topic: PicoScope 6-3-55
Replies: 3
Views: 1940

Re: PicoScope 6-3-55

Hi JC,

Good to see that Picoscope 6 R6.4.10 has solved you issues, please do keep an eye on PicoLabs. Pico Labs is a showcase of our latest software releases.

Kindest regards,

Stuart
Technical Specialist
by Pico Stuart
Wed Apr 07, 2010 1:50 pm
Forum: USB PC Oscilloscopes
Topic: Picoscope 3205 hangs up
Replies: 1
Views: 1445

Re: Picoscope 3205 hangs up

Hi Greg, Thankyou for sending in your observations with regard to your 3205. Please be aware that you scope will function in the later release of the software. If convenient please proceed to the following link and download the latest version of software. http://www.picotech.com/software.html The 32...
by Pico Stuart
Wed Jan 27, 2010 5:56 pm
Forum: Serial Port Products
Topic: ADC 16 USB to serial problems
Replies: 2
Views: 5141

Re: ADC 16 USB to serial problems

Hi Sally, To ensure correct functionality please connect the converter only to the PC first ensuring driver is installed. Look in device manager under Ports (Comm & LPT) windows should recognise this as a 'serial on USB port' . Windows will then assign this a comm port number. If you connect the con...
by Pico Stuart
Mon Jan 04, 2010 11:14 am
Forum: PicoScope 5
Topic: Picoscope crashes when selecting copy as text
Replies: 4
Views: 8429

Re: Picoscope crashes when selecting copy as text

Dear Seuchato, Sorry to see you have encountered these difficulties. I have been able to save your files as .txt using 5.20.9 with an ADC 216 Running. Please proceed to the following link and follow the instuctions accordingly. http://www.picotech.com/software.html In addition to this please find at...
by Pico Stuart
Tue Dec 29, 2009 9:08 pm
Forum: Parallel Port Products
Topic: unstable PH readings
Replies: 13
Views: 9252

Re: unstable PH readings

Hi Gtipc, Thankyou for your observations with regard to your Ph measurements. Looking through your reports looks like you need some filtering on your pump supply. It may be worth looking at some form of isolation transformer. This could possibly be a solution to your interference. Hope that helps. K...
by Pico Stuart
Tue Aug 04, 2009 2:23 pm
Forum: Pre-Sales Advice
Topic: anti alias filter
Replies: 5
Views: 5255

There is no anti-aliasing filter in the hardware, however there is filtering available in the software. Filter details The lowpass filtering algorithm is chosen according to the relationship between the selected cut-off frequency (fC) and the sampling rate (fS), as shown in the following description...
by Pico Stuart
Tue Jul 28, 2009 3:14 pm
Forum: USB PC Data Loggers
Topic: ADC-20 not recognised by Pico
Replies: 1
Views: 2477

ADC-20

Hi Jeremy, Sorry you are having issues with connecting your ADC-20. In order to help me resove this issue please could you try the following:- 1) Open up PicoLog then select file/new settings/recording/sampling rate/ in converter details ensure you have selected ADC20/24 this should enable the ADC 2...
by Pico Stuart
Mon Jul 27, 2009 12:30 pm
Forum: Parallel Port Products
Topic: XY Oscilloscope does not work!
Replies: 4
Views: 3606

X-Y SCOPE CONFIGURATION

Hi There, Please refer to section 5.8 of the attached manual PicoScope 5. (x-y scope configuration) The XY Timebase controls the time interval for each scan. Ideally, it should be a little longer than a single cycle of the signal that you wish to observe. The time per scan field sets the time interv...
by Pico Stuart
Fri Jul 17, 2009 11:16 am
Forum: Parallel Port Products
Topic: XY Oscilloscope does not work!
Replies: 4
Views: 3606

Hi There, ref: ADC 212 Please could you let me know your current settings in order to assist you. Just to recap please try the following if you haven't already. In order to set the X-Y scope view function. Select view/new XY scope this should enable the x-y function. It is possible that your timebas...
by Pico Stuart
Tue Jul 14, 2009 12:43 pm
Forum: Serial Port Products
Topic: TC-08 serial connection
Replies: 4
Views: 7010

Dear Geoff,

Further to your inquiry about the CPU usage percentage. This is normal. The usage is affected by sample rate and is PC specification dependant.

The lower the sample rate the lower the CPU usage. Unfortunately there is no work around for this.

Kindest regards,

Stuart
by Pico Stuart
Tue Jul 14, 2009 11:11 am
Forum: USB PC Oscilloscopes
Topic: Two AWG channels (synchronization issues?)
Replies: 2
Views: 2348

Dear Andrea,

Io order to fully undertand your query please could you provide me with some additional information. Please could you define your superimposed slow transients and your synchronization definitions.

Kindest regards
Stuart
by Pico Stuart
Wed Jul 08, 2009 12:05 pm
Forum: USB PC Data Loggers
Topic: Humidiprobe "unplug - replug"
Replies: 3
Views: 3123

Hi Robert, Sorry to hear that you are struggling to activate humiprobe after hibernation/reboot. I have tried to simulate your fault and can replicate this after hibernation. What I have discovered is if the unplug or replug device warning appears, I closed picolog and then fully shut down my PC. Af...
by Pico Stuart
Tue Jul 07, 2009 10:55 am
Forum: Getting Started
Topic: Pico pH probe
Replies: 1
Views: 2442

Hi there,

Here is a link to information that should answer your questions in regard to your Ph sensor

http://www.picotech.com/experiments/ph- ... index.html

Kindest regards
Stuart
by Pico Stuart
Tue Jul 07, 2009 9:22 am
Forum: USB PC Data Loggers
Topic: TC-08 USB temperature readings fluctuate
Replies: 17
Views: 12950

Good Morning,

Thankyou for your response. The TC08 cannot be a totally isolated design as there would be no way to power the front end. The power supples are produced via the USB.

Kindest regards
Stuart
by Pico Stuart
Mon Jul 06, 2009 3:04 pm
Forum: USB PC Data Loggers
Topic: TC-08 USB temperature readings fluctuate
Replies: 17
Views: 12950

This problem is almost certainly caused by the thermocouple being used touching something which is at a different potential to the TC-08/PC earth. If this voltage is greater than about +/-7V the channel is pulled out of he common mode range of the TC-08. This causes all channels to read very incorre...