Need to clarify which PicoScopes work with PicoLog
I noticed your website has a big banner saying "PicoLog6 now supports PicoScope oscilloscopes" so I downloaded PicoLog6 and connected my PicoScope 2205MSO. Then I got a message saying this model is not supported! If PicoLog6 doesn't support your full oscilloscope range, I suggest you make it clear o...
- Tue Oct 29, 2019 9:09 pm
- Forum: PicoScope 6 for Windows
- Topic: Persistence mode not working with 2205MSO
- Replies: 0
- Views: 3772
Persistence mode not working with 2205MSO
Persistence mode doesn't seem to work with the 2205MSO. Things are fine in scope mode, but when I click the "persistence mode" button I get a blank screen and the message "Waiting for data..." which persists indefinitely. Sometimes attempting to go back to scope mode crashes PicoScope6. PicoScope® 6...
- Sun Jun 17, 2018 8:16 am
- Forum: PicoScope 6 for Windows
- Topic: Calculating time difference between two signals
- Replies: 8
- Views: 6114
Re: Calculating time difference between two signals
I was also looking for the ability to measure the time from an edge on channel A to an edge on channel B, and was disappointed to find that such a basic feature was missing. More specifically, it's the ability to gather stats on the time that would be useful, for example to evaluate the jitter of a ...
- Mon Jan 15, 2018 10:46 pm
- Forum: PicoScope 6 for Windows
- Topic: High pulse width measurement inaccuracy
- Replies: 18
- Views: 7420
Re: High pulse width measurement inaccuracy
Martyn,
Any progress? Is it going to be fixed?
Rgds, Ben
Any progress? Is it going to be fixed?
Rgds, Ben
- Mon Nov 27, 2017 3:09 pm
- Forum: PicoScope 6 for Windows
- Topic: High pulse width measurement inaccuracy
- Replies: 18
- Views: 7420
Re: High pulse width measurement inaccuracy
Yes, you're right, it's not really a problem with pulse width measurement - it looks more like a missing samples / incorrect timebase kind of thing. In retrospect my choice of subject description wasn't very good. Unfortunately it's the best PC I have available, so trying a faster one isn't an optio...
- Sun Nov 26, 2017 6:03 pm
- Forum: PicoScope 6 for Windows
- Topic: High pulse width measurement inaccuracy
- Replies: 18
- Views: 7420
Re: High pulse width measurement inaccuracy
Oops, that last link probably wasn't public... try this one:
https://drive.google.com/drive/folders/ ... sp=sharing
https://drive.google.com/drive/folders/ ... sp=sharing
- Sun Nov 26, 2017 5:59 pm
- Forum: PicoScope 6 for Windows
- Topic: High pulse width measurement inaccuracy
- Replies: 18
- Views: 7420
Re: High pulse width measurement inaccuracy
I wasn't able to attach the files as they were rejected as too large (sigh). So I have put them on Google drive:
https://drive.google.com/open?id=1Jleod ... taFL-hPHPx
Rgds,
Ben
https://drive.google.com/open?id=1Jleod ... taFL-hPHPx
Rgds,
Ben
- Sun Nov 26, 2017 5:50 pm
- Forum: PicoScope 6 for Windows
- Topic: High pulse width measurement inaccuracy
- Replies: 18
- Views: 7420
Re: High pulse width measurement inaccuracy
Martyn, I re-ran it at 100ms/div (m_100ms_div.psdata) and 200ms/div (m_200ms_div.psdata). As before, the results were correct for 100ms/div but wrong for 200ms/div. I turned off the digital channels and tried again at 200ms/div (m_200ms_div_digital_off.psdata). This does seem to fix the problem. I'v...
- Tue Nov 21, 2017 12:46 pm
- Forum: PicoScope 6 for Windows
- Topic: High pulse width measurement inaccuracy
- Replies: 18
- Views: 7420
Re: High pulse width measurement inaccuracy
Martyn, PicoScope® 6 - PC Oscilloscope software version: 6.12.9.2917 Copyright © 1995-2017, Pico Technology Ltd. Model: PicoScope 2205 MSO Serial Number: CZ114/032 USB Version: 2.0 Calibration Date: 30 May 2014 Hardware Version: 1 Driver Version: 1.3.0.14 Firmware Version: 1.3.3.0 / 0.2.48.0 Thanks,...
- Mon Nov 20, 2017 9:01 pm
- Forum: PicoScope 6 for Windows
- Topic: High pulse width measurement inaccuracy
- Replies: 18
- Views: 7420
Re: High pulse width measurement inaccuracy
Bennog,
Thank you for suggesting that topic, but no, it quite obviously does not explain the differences I am seeing.
Ben
Thank you for suggesting that topic, but no, it quite obviously does not explain the differences I am seeing.
Ben
- Mon Nov 20, 2017 8:57 pm
- Forum: PicoScope 6 for Windows
- Topic: High pulse width measurement inaccuracy
- Replies: 18
- Views: 7420
Re: High pulse width measurement inaccuracy
Martyn,
Yes, if I change the slow sampling transition to 500ms/div, then the measurements are good at 200ms/div. So it appears there is a bug which only affects slow sampling mode - do you agree?
Thanks,
Ben
Yes, if I change the slow sampling transition to 500ms/div, then the measurements are good at 200ms/div. So it appears there is a bug which only affects slow sampling mode - do you agree?
Thanks,
Ben
- Mon Nov 20, 2017 3:37 pm
- Forum: PicoScope 6 for Windows
- Topic: High pulse width measurement inaccuracy
- Replies: 18
- Views: 7420
Re: High pulse width measurement inaccuracy
Bennog, Thanks for pointing this out, but I was already aware of it. The thing is, I'm seeing good accuracy at the lower sampling rate, and lousy accuracy at the higher sampling rate. At 12.27KS/sec you'd expect to measure a 500ms interval to about 0.02% accuracy. At 2.5MS/sec, you'd obviously expec...
- Thu Nov 16, 2017 9:21 pm
- Forum: PicoScope 6 for Windows
- Topic: High pulse width measurement inaccuracy
- Replies: 18
- Views: 7420
High pulse width measurement inaccuracy
I have a circuit which generates an accurate pulse 500ms wide. I use the PicoScope 6 measurement function to measure the high pulse width between rulers. When the timebase is set to 100ms/div, I get sensible results: Min=500.1ms; Max=500.1ms; Average=500.1ms; sigma=24.46ns If I change the timebase t...